SABRATEK 3030 IV Pump

The Sabratek 3030 infusion pump provides accurate, safe, and reliable volumetric delivery and can be used in both arterial and venous applications.

  • The pump is easy to program and use. You simply agree or disagree with the pump's prompts by using the YES and the NO keys.
  • The pump's Lockout mode makes it tamper-resistant by requiring a software access code to be keyed in prior to any parameter changes. This limits patient programming and profile selection while maintaining optimum compliance.
  • The Sabratek pump offers four delivery profiles:
    1. Continuous profile - a continuous delivery of a single flow rate.
    2. Auto-Ramp profile - automatic, linear ramping (tapering) capabilities.
    3. Intermittent profile - delivery of a preset number of doses at regular intervals.
    4. 15 Periods profile - sequential delivery of up to fifteen different rates and volumes.
  • Secondary (Piggyback) infusion can be delivered before or during the primary infusion. The pump automatically switches to delivering the primary infusion's rate when the Piggyback volume is complete.
  • Titration allows the delivery rate and/or the volume to be infused to be changed in Continuous and in Piggyback profiles.
  • The pump provides a choice of entering either a delivery rate or a delivery time for an infusion.
  • You may delay the starting time of any profile (including Piggyback).
  • Programmed infusion settings are recalled by the pump indefinitely, or until they are changed by the operator. There is no need to reprogram the parameters before each use.
  • Turning the pump off during an infusion requires two separate keys to be pressed (decreasing the chances of an accidental interruption of an infusion).
  • If the pump is turned off before an infusion is completed, the rate, volume-to-be-infused, and volume infused will be retained in the pump's memory,. When the pump is restarted, the infusion may be resumed exactly from where it was stopped.
  • The audible alarm cannot be completely disabled. Quiet operation.
  • Pump information can be displayed in either English or Spanish (other languages are optional).
  • Pump displays the time remaining until the end of an infusion.
  • The volume infused display can be cleared to zero
  • The pump displays a flashing flow indicator arrow while the pump is infusing. It is displayed between the rate and the volume infused displays and flashes every one second

SPECIFICATIONS

Flow Rate Range 1 ml/hr to 999 ml/hr in 1 ml/hr increments
Volume Limit Range 1 ml to 9999 ml in 1 ml increments
Accuracy +/-5% with standard, soft sets
Pumping System Linear peristaltic
Weight approximately 10.5 lbs (4.5 kg).
Dimensions 8" (20.5 cm)H x 5.5" (14 cm)W x 6.25" (16 cm)D
Power Requirements 110V AC, 10 Watts, 50/60 Hz
Power Operating Range 100 to 130V AC
Over Current Protection 0.25 Amp AC line fuse
Battery 12 volt, rechargeable
Battery Operating Range 6 to 7 hrs at 125 ml/hr (run-time depends on the state of the battery charge)
Battery Recharge Time approximately 24 hours from empty battery
Occlusion Alarm LO: 8 psi, +/- 5 psi, HI: 20 psi, +/- 5 psi
KVO Rate 1 ml/hr; 1 to 9 ml/hr in Intermittent delivery
Power Cable approximately 10' long, hospital grade
   
ADMINISTRATION SETS Non-proprietary, standard, soft, straight-line set