IVAC MedSystems III
Multi-Channel IV Pump

  • Three independent fluid delivery systems in the space of one
  • Compact size:
    • reduces bedside clutter
    • simplifies patient transport
  • Easy to set up and use, yet provides advanced features
  • Accurate delivery of a variety of fluids
  • Accommodates assorted container types
  • Multiple delivery methods: Intravenous/Intra-arterial/Subcutaneous/Epidural
  • Uses administration sets that provides free flow protection . Six available Device Types with configurable parameters (maximum and minimum rates, maximum volumes, baseline and maximum pressures, and air-in-line thresholds) to achieve specific clinical applications:
    • General Purpose
    • Neonatal
    • Controller Pressure
    • Operating Room
    • General Purpose II
    • Operating Room II
  • Displays infusion status for rate, volume remaining and volume infused.
  • Infusions can be programmed to deliver at a specified rate or over a specified period of time.
  • Secondary mode allows fluids and medications to be delivered at two different rates, sequentially.
  • Dose Rate Calculation (DRC) feature performs the volumetric rate and/or dose rate calculations.
  • With DRC activated, displays infusion status for rate, dosing regimen and drug name.
  • Communications Protocol allows clinical monitoring, instrument configuration and maintenance.
  • Field Maintenance Software (FMS) available for Biomed to configure, service and troubleshoot the pump.

FEATURES

Multi-Channel Fluid Delivery System
The instrument combines three independent infusion channels in an unparalleled small size.

Lightweight/portable
The pump with pole clamp weighs just over 5 pounds and is easy to transport.

Unique rotating pole clamp
The Pump may be attached to a variety of surfaces.

Dose Rate Calculator (DRC)
The Pump calculates a volumetric or dose rate based on values entered for patient weight, drug concentration (drug amount and diluent volume) and dosing parameters.

Six Device Types Available
General Purpose, Neonatal, Controller Pressure, Operating Room, General Purpose II, and Operating Room II.

Free-Flow Protection
The IVAC 28 and 25 Series Administration Sets contain a cassette that provides protection from free-flow conditions. To remove the cassette from the pump, the cassette's slide clamp is pulled to full extension, occluding the tubing and preventing fluid from flowing.

Monitoring System
The instrument continuously monitors pump conditions and alerts with adjustable audio tones and visual messages.

Data Monitoring
The pump can be configured to communicate with a remote computer, such as a centralized patient monitoring nurses station. The COMM receptacle is compatible with RS-232 cabling. A communications manual that describes the programming and hardware involved is available.

Field Maintenance Software (FMS)
The pump can be modified to accommodate specialized clinical applications. The Device Type parameters, occlusion limit, and air-in-line threshold can be reconfigured with the optional FMS software.

Secondary Mode
Allows the user to program two different rates of infusion to run sequentially.

Full Range of Delivery Rates
Rates from 0.1 to 999 milliliters an hour.

Battery Capacity
A fully-charged battery provides 6 to 8 hours of operating time with rates of 125 ml/h per channel.


SPECIFICATIONS

Standards UL 544
CSA C22.0, No 125
Case Material Impact resistant polycarbonate/ABS alloy
Dimensions Height 7.875 inches (20.00 centimeters)
Width 6 inches 815.24 centimeters)
Depth 2.10 inches (5.33 centimeters)
Weight Approximately 5.1 lbs. (2.32 kg) includes Pole Clamp
Air-In-Line (Default) 500 µl except for Neonatal which is 50µl
Occlusion Pressure (Default) 15 psi except for Controller pressure device which is 3 ft H20
Operating Temperature 50- 104º Fahrenheit (10º - 40º Celsius)
Storage Temperature <95º Fahrenheit (<35º Celsius) for optimum battery life
Maximum Storage Temperature 131º Fahrenheit (55º Celsius) Temperature
Rate Range 0.1 - 999 milliliter per hour (each channel)
Volume Range 0.1 - 999 milliliter (each channel)
KVO Rate Range 0.1 - 20.0 milliliter per hour
System Accurac ±5%
Power Consumption 6 watts AC power. Use only IVAC MedSystem III AC Adapter, Model 1550
Batteries

Main - Rechargeable NiCD Battry Pack Memory
Back-up - Nonrechargeable Lithium

Battery charge

A fully charged battery has a minimum of 6 hours running time with all channels running at 125 milliliter per hour and backlighting usage of 2 minutes per hour.

The main battery retains 80% of its capacity after 500 charging Cycles, and retains 90% of its capacity after 3 months of Continuous AC charging.

AC Adapter & Cord Length Model 1550, 8.5 VDC @ 750 mA & 8.5 feet (2.59 meters)
AC Adapter Connector 4 pin locking connector is standard on Model 2863. Detachable Connector is standard on Model 2860
Fuses 3 amp fast-blow internal
Ground Continuity Maximum 0.1 ohm
Leakage Current Maximum 100 microamps
   
ADMINISTRATION SETS Use only IVAC MedSystem III Administration Sets