IVAC
7100 & 7200 The IVAC Signature Edition Infusion System includes the Model 7100 and Model 7200 Volumetric Infusion Pumps and AccuSLIDE Flow Regulator administration sets. This document provides directions for use for both model pumps.
The pumps may be used to deliver a wide variety of fluids over a broad range of infusion rates from multiple fluid container types. |
The system is designed to be used in many areas of patient care, including: |
|
|
The pumps' applications include: |
|
|
The Signature Edition Infusion system uses a wide variety of AccuSlide Flow Regulator Administration sets. The IVAC 72 series sets are designed for use within the pumps as well as for gravity-flow, stand-alone use. The unique, patented AccuSlide flow Regulator has an integral flow control device that minimizes accidental free-flow when the set is removed from the pump and provides accurate rate control during gravity administration. |
SPECIFICATIONS
Flow Rate Range | 0.1 to 999.9 ml/hr
in 0.1 ml/hr increments (primary) 1.1 to 270.0 ml/hr in 0.1 ml/hr increments (secondary) |
Volume To Be Infused Range | .1 to 9999.9 ml
in 0.1 ml increments (primary) .1 to 999.9 ml in 0.1 ml increments (secondary) |
Volume Infused Range | 0.0 to 9999.9 ml in 0.1 ml increments |
Alarms |
Air in Line |
Secondary | Set Out |
Set Up Time Exceeded | |
Dimensions | |
7100 | Width 7.7 in/19.3 cm |
Height 8.6 in/21.8 cm | |
Depth* 5.0 in/12.7 cm | |
Weight** 6.85 lbs./3.11 kg | |
Power Cord 10 ft/3 m | |
7200 | Width 10.5 in/26.7 cm |
Height 8.6 in/21.8 cm | |
Depth* 5.0 in/12.7 cm | |
Weight** 8.4 lbs./3.7 kg | |
Power Cord 10 ft/3m | |
*without pole clamp | |
Case | Impact resistant plastic |
Power Requirements | 100- 120 V~,50/60 Hz, 0.5A,3-wire grounded system (40 watts Max) |
Ground Current Leakage | Tested to UL Standard 544 and CSA C22.2 No. 125 for medical and dental equipment. |
Battery | Rechargeable nickel cadmium. With a new, fully charged battery, the pump will operate for 4 hours nominal at 100 ml/hr for a two channel instrument operating on both channels simultaneously. |
Environmental Conditions | |
Operating | Temperature Range
10ºC to 40ºC (50ºF to 104ºF) |
Relative Humidity 15 to 95% Non-condensing | |
Atmospheric Pressure 632 to 1031 mbar | |
Storage | Temperature Range -40ºC to 60ºC (-40ºF to 122ºF) |
Relative Humidity 5 to 95% | |
Atmospheric Pressure 632 to 1031 mbar | |
ADMINISTRATION SETS | Use only IVAC 72 series administration sets. |