Horizon NXT
Positive pressure is frequently a necessity but clinical experience shows that high pressure limits can increase the severity of an infiltration without causing an alarm. Because there is a need to control pump pressure settings, the Horizon NXT has user-selectable occlusion pressure limit settings which start at 75 mmHg and extend to 500 mmHg. At rates of 400 ml/hr and higher, the pressure setting is automatically increased to 400 mmHg. For epidural infusions which require higher pressures, the pump has an extended occlusion pressure setting of 750 mmHg. Because some infusions require multiple flow rates during the delivery cycle, the Horizon NXT is equipped with two automatic multiple delivery rate modes. The Profile Delivery mode incorporates a standard ramped delivery regimen, and the Programmable Delivery mode can be set up to sequentially deliver up to nine user-selected volumes of fluid each at a user-selected rate. Both modes are designed for ease of use and can be adapted to home-care settings. The Horizon NXT also incorporates a Dose/Rate mode which automatically calculates the rate when dose information is entered, or the dose when rate information is entered. The pump is equipped with distinct audible and visual alarm signals to indicate KVO, low battery, and actual alarm conditions. Epidural
Administration Caution: Whenever the pump is being used for epidural infusions, the pump's secondary infusion features should not be used. Blood
Infusion q(x)º |
Specifications
Mechanism | Volumetric (non Drop-Counting), Positive Pressure Displacement Reservoir |
Weight | Approximately 11.9 lb. (5.4 Kg) |
Power Requirements | 90-132 VAC, 30 Watt, 47/63 Hx excluding accessories (domestic), 180 - 264 VAC, 30 Watts, 47/63 Hz excluding accessories (International) |
Grounding Resistance | Meets UL Standard 544 |
Leakage Current | Meets or exceeds UL Standard 544, Section 27.2 and ANSI/AAMI Safe Current Limits (July 9, 1985) |
Plug | Hospital Grade (3 pin) |
Pole Size Range | 0.75" to 1.25" |
Battery | 12 Volt sealed lead-acid battery for ambulatory use. Run time is dependent on the state of battery charge and the rate of fluid delivery. |
Rate Range | Standard: (Primary and Piggyback) 0.1 to 999.9 ml/hr Micro: (Primary and Piggyback) 0.1 to 99.9 ml/hr |
Volume to be Infused | 0.1 to 9999.9 ml |
Occlusion Pressure | 75, 100, 200, 300, 400, 500 and 750 mmHg Standard mode default is 300 mmHg. |
Keep Open Rate | 3 ml/hr for prescribed delivery rates of 3 to 999.9 ml/hr, or the continuation of the prescribed delivery rate for 0.1 to 2.9 ml/hr settings. |
Volume Delivery Accuracy | +/- 5% |
Fluid Types | All standard I.V. Fluids |
Alarms | Air-In-Line, Container Empty, Door Open, Downstream Occlusion, Hold Time Exceeded, Low Battery, Low Flow From Container, Close Roller Clamp, System Error, Upstream Occlusion. |
Air-In-Alarm | 0.09 cc |
Hold Time Exceeded Alarm | Activated after three minutes on hold without user interaction (deactivated if the door of the pump is open) |
Memory | Permanent data retention after the pump is turned off, unless the battery is depleted or disconnected, or unless the data Retention Defaults have been otherwise selected in the Biomed Options. |
Administration Sets | Horizon Pump IV Set, 15 Drops/mL, with check valve and 3 injection Sites |