BAXTER COLLEAGUE IV Pump

Baxter's Colleague 3 Volumetric Infusion Pump is designed to meet the fluid delivery needs of today's evolving healthcare environment. The Colleague 3 Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications.

Fluid delivery applications include:

  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.

The Colleague 3 Volumetric Infusion Pump feature spectrum includes:

  • Micro and Macro Rate Range
  • Basic Delivery Programming
  • Piggybacking Secondary Medications
  • Special Programming functions for Dosing

SPECIFICATIONS

Ac power Requirements 100-120 VAC, 60Hz
Leakage Current Less than 300 u amps earth leakage (tested per UL 2601)
External Fuses 1.6 amp Type T (time delay) 250V
Power Cord 2.9 m (9 feet) long with integrally molded hospital grade plug
Battery Supply System Colleague 3 Pump - 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries) 12 hours to recharge to 80% of capacity Internal charge system recharges batteries whenever the device is connected to an AC outlet
Range of Programmable Flow Rates Primary Infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments 1 to 1200 mL/hr in 1 mL/hr increments Piggyback Infusion on each pump channel 0.1. to 99.9 mL/hr in 0.1 mL/hr increments 1 to 500 mL/hr in 1 mL/hr increments Note: Rate limits can be configured for values less Than those stated above.
Volume to be Infused 0.1 - 99.9 mL in 0.1 mL increments 1- 9999 mL mL/hr in 1 mL increments
Note: Volume to be Infused limits can be configured for values less than those stated above
KVO 0.1 - 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate whichever is less.
Priming rate 500 mL/hr (to be used only when not connected to patient)
Advance Air Rate to Advanced an Air Bubbles At the same rate programmed for the current primary or piggyback infusion.
Air Bubble Setting Measured accumulations of approximately
25 microliters
50 microliters
100 microliters 150 microliters
The Air Bubble Setting is a configurable option.
Auto Restart Allows the pump channel to automatically restart if an occlusion is relieved within one minute after detection. This configuration option allows 0 to 9 restarts before manual intervention is required.
Nominal Downstream Alarm sensitivity is a configurable option. Rate rangein mL/hr
<21 21- 200 >200
2 psig 4 psig 6 psig Minimum
5 psig 8 psig 11 psig Moderate
9 psig 12 psig 15 psig Maximum
Device Size Approximately 13.88" H x 7.75" W x 8.00" D
(353 x 197 x 203 mm)
Not including clamp knob or power cord
Device weight Approximately 17.5 lbs (7.9 kg)
Environmental Operating Limits 59º F - 100º F (15º C - 38º C)
Environmental Storage and Transport Limits -20º F - 135º F (-29ºC - 57º C), 10-100% Reliable Humidity including condensation, 500-1060 hPa Barometric Pressure
Accessories Configuration Transfer Cable (Product code 2M8155)
Ground Impedance Less than 0.2 ohms (tested per UL-2601)
Solution Container Height Pump will operate at all programmable flow rates with a solution container height ranging from -36 to +48 inches. Container height is measured from the pump channel selected for use.
Interfaces Attachment to a remote nurse call system A variety of systems exist which enable alert and alarm conditions to be transmitted to a remote location via the communications port on the rear panel assembly of the pump. See the service manual for additional information.
 
ADMINISTRATION SETS Baxter sets equipped with keyed On/Off clamps. 2C5331, 2C5417, 1C8160, 2C5431, 2C5418
 