BAXTER/TRAVENOL
FLO-GARD 6200 The Flo-Gard 6200 Volumetric Infusion Pump is an electromechanical device used for the intravenous infusion of liquids at user-selected rates. The pump operates on standard 115 VAC 60 Hz electrical power, or on its self-contained rechargeable battery. It is portable, has a panel lock-out feature to discourage patient tampering, and permits infusion of primary and secondary medication programs. The secondary program automatically switches over to the primary program when secondary infusion is complete (automatic piggybacking). The pump is designed for use with Baxter standard administration sets which contain an "s" as the last character of the code number, for example, 2C5537s. When infusing solutions through a central venous catheter, sets with a luer lock adapter should be used. Sets with a Flashball device are not recommended in these applications. The primary rate of infusion is selectable from 1 to 999 mL/hr. The secondary rate is selectable from 1 to 999 mL/hr. The volume to be infused (VTBI) is selectable between 1 and 9999 mL. The total volumes infused from primary and secondary programs are added together and accumulated and can be displayed on demand. The primary and secondary VTBIs are independently decremented and displayed. Upon completion of the primary VTBI, the pump automatically switches to a keep vein open (KVO) rate. If the pump is started on a secondary rate and VTBI, the pump will change to the primary rate when the secondary program is completed. The pump may be stopped at any time by pressing the STOP key, unless in the Lockout mode. |
PHYSICAL SPECIFICATIONS
| Approximate Dimensions | 24 cm W x 16 cm D x 27 cm H (9.5" W x 6" D x 11" H) |
| Internal Battery | Rechargeable, 8-volt sealed lead-acid |
| Battery Charge Time | 24 hours to completely recharge |
| Power Cord | Replaceable 2.4 m (8 ft) long with hospital-grade plug |
| Fuses | Type 0.5A, 250V, SB 6.35 mm |
| Leakage Current | Less than 20 microamps (per UL 544) |
| Outer Casing | Molded plastic |
| Operating Environment | 10º-40º C (50º - 104º F) |
| Approximate Net Weight | 6.7 kg (14 lb) |
| Power Requirements | 115 VAC, 60 HZ, 10 Watts |
PERFORMANCE SPECIFICATIONS
| Flow Rate Range | 1-999 mL/hr in 1 mL/hr increments |
| Volume Range | 1-9999 mL in both primary and secondary |
| KVO Rate Ranges | 1-5 mL/hr |
| Battery Life | Approximately 5 hours on fully charged batteries at 125 mL/hr |
| Air In Line Detection | Air bubbles 1/3 in (.05 mL) or longer |
| Gravity Free Flow Protection | An automatic safety clamp occludes tubing when the pump door is opened |
| Administration Set | Standard administration sets from Baxter |