ABBOTT
Lifecare 5000 Plum
|
Specifications
Dimensions: | Approximately 18 x 23 x 23 cm (7 x 9 x 9 in), excluding pole clamp protrusion and power cord storage. |
Weight: | Approximately 6.0 Kg (13 lb), with battery. |
Casing: | High-impact plastic |
Power requirements: | 110-120 VAC, 50/60 Hz, 30 W. |
Power Cord: | Hospital-grade AC cord. 8 ft long, with transparent Plug and retainer plate. |
Fuses: | 0.5 A, 250 V, Slow Blowing. |
Battery: | One sealed, lead-acid,
rechargeable 8 V battery, Internal to device. Accessible for ease of
field replacement, with color-coded leads and polarized connector. Battery life (new batteries, full charge, at 20 C) approximately 500 mL total volume delivered or six hours of operation, whichever occurs first |
Delivery Rate Range
Micro Mode: | 0.1 to 99.9 mL/hr (in o.1 mL. Increments) The total of the primary rate plus the secondary rate Cannot exceed 99.9 mL/hr in Micro Mode. Note: While using the concurrent delivery feature in the Micro Mode, neither the primary rate nor the secondary rate can be less than 0.5 mL/hr. |
Macro Mode: | 1 to 999 mL/hr (in 1 mL increments. the total of the primary rate plus the secondary rate cannot exceed 700 mL/)in 1 mL/hr in Macro Mode |
Dose Limit Range
Micro Mode: | 0.1 to 999 mL (in 0.1 mL increments). |
Macro Mode: | 1 to 9999 mL (in 1 mL increments). |
Operating Backpressure
-2 to 10 psig. The maximum pressure is user-selectable from 1 to 10 psig, via the front panel touchswitches. |
|
Secondary Doses: |
For dual-channel delivery, a single dose of a secondary fluid may be administered. For multidose delivery, 1 to 24 doses of a secondary fluid may be administered at intervals from 15 minutes to 24 hours. |
Occlusion Alarm
Distal: |
The Distal Occlusion alarm sounds within two pumping cycles after the distal set tubing or set outlet fitting becomes occluded. |
Proximal: |
The Proximal Occlusion alarm sounds within five pumping cycles if the tubing proximal to the cassette becomes occluded. |
Air-in-alarm: | The Stopped Air In Distal Line alarm will sound if a bubble 100 uL or larger passes the distal air-in-line sensors. (alarm may actually sound at detection of a bubble as small as 50 uL.) |
Proximal: | The Stopped Air In Proximal Line alarm will sound if approximately 1200 uL or more passes through the proximal air-in-line sensors. |